You are here:

DS400 GfG Docking Station

Get A Quote

The ultimate DS400 GfG Docking Station for instrument maintenance for GfG’s G450 and G460 gas detectors.

For best pricing, please call 1-800-952-3293 or email sales@afcintl.com

Description

Description

The ultimate DS400 GfG Docking Station for instrument maintenance for GfG’s G450 and G460 gas detectors.

The operation of the DS400 docking station couldn’t be any easier. Simply put the G450 or G460 in the dock and the system will take over. The DS400 automatically performs a full functional test including a fresh air calibration, bump test, extensive sensor tests, alarm tests and optionally performs an automatic AutoCal calibration. The DS400 is designed to go beyond the traditional basic docking station test and includes sensor diagnostics tests such as T50 and T90. After all, what good is a sensor that can respond to gas when it takes too long to get there? With the DS400 you can have confidence with this added level of safety.

After all tests have been completed the DS400 will automatically start the charging process of the detector or you may remove and begin using it. In case the detector’s battery condition is insufficient to complete a test the DS400 will charge the instrument before performing each test.

A good instrument gas detection maintenance program is not complete without equally good records keeping. Whether you choose to download or maintain the records on a PC or network or your preference is to leave it to the docking station, the DS400 is the right choice. The DS400 can retain as many as 2 million complete diagnostics in memory with maximum memory installed. That is over 50 years of data even if testing 100 detectors every single day.

The DS400 is designed to reduce the cost of ownership every step of the way. The built in pump has variable flow control to reduce the amount of gas consumed during the test. The DS400 is even smart enough to take advantage of the gas remaining in the tubing after the previous test was completed.

The DS400 software includes a module to quickly print official calibration or bump test certificate. simply download the records from the DS400 and start the Certificate Generator.

Features

  • Fully automatic functional bump test and calibration
  • Automatic test of alarms, charging & record keeping
  • Simple pas/fail indication
  • Saves last calibration and bump test records in detector for easy retrieval of maintenance record in the field
  • No computer necessary
  • Up to 2,000,000 records in memory
  • Cascade multiple docking stations
  • Save time, gas and efforts for lower cost of ownership
  • Print official calibration certificates

Specifications

Test Protocol Fresh Air: Automatic calibration
Gas Response: Automatic bump test, automatic or selective calibration
Response Time: T50 & T90 tests
Alarm Activation: Time to activate alarm 1 & alarm 2
Visual Alarm: Automatic LED tes
Audible Alarm: Automatic horn test
Interface G450 or G460 keypad
Display G450 or G460 graphic display
Time Stamp Automatic/Real time clock
Inlets 1 gas and 1 fresh air
Outlets 1 gas exhaust
Flow Rate 0.5 LPM, variable flow
Temperature Range 50-90°F
Power 12 VDC wall cube, 12 VDC vehicle power (optional)
Size 5.1″ x 9″ x 4.7″
Weight 28.5 ounces
Warranty 1 year

Ordering Information

Each DS400 includes software, USB computer cable, power supply, 1 GB SD memory card, download software, certificate generator. For pricing, please call AFC International at 1-800-952-3293, 219-987-6825 or fax 219-987-6826 or email sales@afcintl.com.

Description
Catalog No.
Docking Station, instrument only 1450-401
Docking Station, instrument and pump 1450-402
Docking Station, includes demand flow regulator, instrument only 1450-401R
Docking Station, includes demand flow regulator, instrument and pump 1450-402R
Four Inlet Multi-gas Docking Station, instrument only 1450-404
Four Inlet Multi-gas Docking Station, instrument and pump 1450-405
Demand Flow Regulator 1419-216