You are here:

Hydrogen Cyanide Replacement Sensor for AreaRae / MultiRae / ToxiRae Pro

Ships directly from the manufacturer, shipping times may vary.

$585.00

Description

Description

Genuine Hydrogen Cyanide Replacement Sensor for AreaRae / MultiRae / ToxiRae Pro gas detectors from Rae Systems by Honeywell.

All specifications have been verified under the following environmental conditions:

  • Temperature: 68°F
  • Relative Humidity (non-condensing): 50%
  • Ambient Pressure: 1 atm (1,013 mbar)

Refer to the Technical Note TN-114 for a glossary of terms used on this page

Additional information

Weight .5 lbs

Specifications

Sensor Type Electrochemical
Range 0-50 ppm
Resolution 1 ppm
Maximum Overload 100 ppm
Drift <2% signal/month
Bias/Equilibration No bias/10 minutes after installation
Storage Life 6 months in sealed container
Operating Life 2 years in air
Calibration Gas 10 ppm Hydrogen Cyanide in Nitrogen
Temperature Range -4°F to 122°F
Storage Temperature Range 32°F to 68°F
Pressure Range Atmospheric ±10%
Operating Humidity 15 to 90% rH non-condensing
Warranty 1 year from date of shipment

Cross Sensitivities

Gas
Concentration
Response
Carbon Monoxide
300 ppm
15 ppm
Ethylene 100 ppm 25 ppm
Hydrogen
200 ppm
0 ppm
Hydrogen Sulfide
15 ppm
90 ppm1
Nitric Oxide
35 ppm
-28 to 0 ppm2
Nitrogen Dioxide 5 ppm -20 t0 10 ppm2
Sulfur Dioxide
20 ppm
40 to 75 ppm

1 – Due to a very high cross sensitivity to Hydrogen Sulfide, this sensor is unsuitable for use in atmospheres that contain Hydrogen Sulfide
2 – CAUTION! Negative cross sensitivities may cause the sensor to produce lower readings than the true concentration of gas in ambient air.

Downloads

Technical Note: TN-114 – Sensor Specifications and Cross Sensitivities for Hydrogen Cyanide Replacement Sensor

Disclaimer: Technical Note TN-114 presents specifications, cross sensitivities and calibration information on select Rae Systems by Honeywell sensors. All specifications presented in this Technical Note reflect the performance of standalone sensors. For instrument specifications, please refer to product datasheets and manuals. Actual sensor characteristics may differ when the sensor is installed in different instruments. As sensor performance may change over time, specifications provided are for brand new sensors.

Product Enquiry