The CAL 2000 Gas Generator features field replaceable, electrochemical generating cells provide a calibration standard for accurately testing chlorine, chlorine dioxide, hydrogen, hydrogen cyanide and hydrogen sulfide gas sensors.

Sources are available here.

For best pricing, please call 1-800-952-3293 or email sales@afcintl.com

$2,253.00

Description

Description

The CAL 2000 Gas Generator features field replaceable, electrochemical generating cells provide a calibration standard for accurately testing chlorine, chlorine dioxide, hydrogen, hydrogen cyanide and hydrogen sulfide gas sensors.

The CAL 2000 is microprocessor controlled and has an LCD display allowing field adjustable ppm and flow rate. The generating cells are interchangeable and field replaceable. Best of all, the CAL 2000 provides up to 50 times as much calibration gas as a disposable cylinder at approximately the same cost. The electrochemical generating cells do not degrade over time.

The Cal 2000 gas generator has a built in mass flow sensor that automatically compensates for altitude and temperature. The mass flow sensor combined with the instruments internal pump, provide extreme flexibility for calibrating both diffusion and sample draw detection systems.

Additional information

Weight 4 lbs
Cal 2000

Cal 2000 Gas Generator

Features

  • Demand flow capable
  • A compact hand held size
  • Automatic purge on shut down
  • Low power consumption
  • Alkaline battery powered
  • Adjustable ppm and flow rate
  • Interchangeable calibration gases
  • NIST traceable certificate
  • CSA NRTL/C Exia T-6 approved intrinsically safe with CE mark
  • ATEX II3G EEx nL IIB T4 approved

Specifications

Hydrogen Sulfide
0.5-50 ppm
Hydrogen Sulfide Micro-Cell
0.05-1 ppm
Chlorine
0.5-50 ppm
Chlorine Micro-Cell
0.05-1 ppm
Chlorine Dioxide
0.5-5.0 ppm
Hydrogen Cyanide
0.5-50 ppm
Hydrogen Cyanide Micro-Cell 0.1-2.0 ppm
Hydrogen
0.5-50 ppm
Air Flow Rate, with internal pump
0.2 to 1 LPM
Sample draw Rate, with pump disabled
0.2 to 5 LPM
Useful Cell Life
10, 50 or 100 hours
Accuracy
±10%
Repeatability
±5%
Warm up Time to 90%
About 5 minutes
Operating Temperature
0 to 50°C
Relative Humidity
0-100%
Battery Power
4 Alkaline C cells
Battery Life
10 hours at 0.5 LPM
Size
8.5″ x 4.25″ x 3″
Weight 3 pounds
Approvals CSA NRT/C Exia T-6
ATEX II3G EEx nL IIB T4

Please Note: For Cal 2000: ppm is variable with flow rate, output given at 0.5 LPM