Description
Genuine Phosphine Replacement Sensor for AreaRae / MultiRae / ToxiRae Pro gas detectors from Rae Systems by Honeywell.
All specifications have been verified under the following environmental conditions:
- Temperature: 68°F
- Relative Humidity (non-condensing): 50%
- Ambient Pressure: 1 atm (1,013 mbar)
Refer to the Technical Note TN-114 for a glossary of terms used on this page
Phosphine Replacement Sensor Specifications
Sensor Type | Electrochemical |
Range | 0-20 ppm |
Resolution | 0.1 ppm |
Drift | <2% signal/month |
Bias/Equilibration | No Bias/10 minutes after installation |
Storage Life | 6 months in sealed container |
Operating Life | 1 year in air |
Calibration Gas | 5 ppm Phosphine in Air |
Temperature Range | -4°F to 122°F |
Storage Temperature Range | 32°F to 68°F |
Pressure Range | Atmospheric ±10% |
Operating Humidity | 15 to 90% rH non-condensing |
Warranty | 1 year from date of shipment |
Cross Sensitivity Data
Gas | ||
Carbon Monoxide | ||
Ethylene | 100 ppm | 0 ppm |
Hydrogen | ||
Hydrogen Sulfide | 12 ppm | |
Ammonia | 50 ppm | 0 ppm |
Sulfur Dioxide | 0.9 ppm |
Downloads
Disclaimer: Technical Note TN-114 presents specifications, cross sensitivities and calibration information on select Rae Systems by Honeywell sensors. All specifications presented in this Technical Note reflect the performance of standalone sensors. For instrument specifications, please refer to product datasheets and manuals. Actual sensor characteristics may differ when the sensor is installed in different instruments. As sensor performance may change over time, specifications provided are for brand new sensors.