Description
Genuine High Range Carbon Monoxide Replacement Sensor for AreaRae / MultiRae / ToxiRae Pro gas detectors from Rae Systems by Honeywell.
All specifications have been verified under the following environmental conditions:
- Temperature: 68°F
- Relative Humidity (non-condensing): 50%
- Ambient Pressure: 1 atm (1,013 mbar)
Refer to the Technical Note TN-114 for a glossary of terms used on this page
Specifications
Sensor Type | Electrochemical |
Range | 0-2,000 ppm |
Resolution | 10 ppm |
Bias/Equilibration | No bias/10 minutes after installation |
Drift | < 2% signal loss/month |
Storage Life | 6 months in sealed container |
Operating Life | 2 years in air |
Calibration Gas | 100 ppm Carbon Monoxide in Air |
Zero Gas | 99.9% Nitrogen |
Temperature Range | -4°F to 122°F |
Storage Temperature Range | 32°F to 68°F |
Pressure Range | Atmospheric ±10% |
Operating Humidity | 15 to 90% rH non-condensing |
Warranty | 2 years from date of shipment |
Cross Sensitivity Data
Gas | ||
Chlorine | ||
Ethanol | 200 ppm | 0 ppm/0 ppm |
Ethylene | ||
Hydrogen | ||
Hydrogen Sulfide (H2S) | 0 ppm/0 ppm | |
Nitric Oxide | 35 ppm | -10 to 0 ppm4/0 ppm |
Nitrogen Dioxide | 0 ppm/0 ppm | |
Sulfur Dioxide | 5 ppm | 0 ppm/0 ppm |
1 – New sensor specs. Used sensors show increasing response to VOCs.
2 – A disk shaped activated carbon filter placed on top of the CO sensor helps reduce the response to VOCs
3 – Not tested (NT)
4 – CAUTION! Negative cross sensitivities may cuase the sensor to produce lower readings than the true concentration of gas in ambient air.
Downloads
Disclaimer: Technical Note TN-114 presents specifications, cross sensitivities and calibration information on select Rae Systems by Honeywell sensors. All specifications presented in this Technical Note reflect the performance of standalone sensors. For instrument specifications, please refer to product datasheets and manuals. Actual sensor characteristics may differ when the sensor is installed in different instruments. As sensor performance may change over time, specifications provided are for brand new sensors.